Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual

8. Change the oscilloscope trigger source from CH2 to CH1.
Trigger
Source . .................... CH1
9. Verify that the Marker2 delay function is operating correctly:
a. Push the Marker2 Delay side button.
b. Continuously change the Marker2 delay from 0 s to 1.5 ns by turning the
general purpose knob.
c. Verify that the Marker2 output delay relative to the Marker1 output and
varies from 0 s to 1.5 ns on the oscilloscope screen.
d. Verify that the Marker2 output delay relative to the Marker1 output is
in 1.05 ns to 1.65 ns at marker2 delay setting is 1.5 ns.
10. Disconnect the oscilloscope.