Report Format
The data format for measurements consists of two
major fields: the measurement units and the measureĆ
ment value(s). See Figure 3Ć3.
WM 0.000E0
XYZ 0.000E0,0.000E0,0.000E0
Measurement
Units
(See Table 3Ć3)
Measurement
Value
Exponential
Multiplier
Figure 3Ć3:ăRSĆ232D Report Format
Measurement unit abbreviations are listed in TableĂ3Ć3.
Values are displayed with four significant digits, and a
oneĆdigit exponential multiplier with an optional minus
sign. For multiple values (for example, the three values
reported for an XYZ reading), commas separate the
values.
Each line is terminated by a carriage return and line
feed (CR/LF).
RSĆ232D Interface
J17 Instruction Manual
3-5
Report Format
The data format for measurements consists of two
major fields: the measurement units and the measureĆ
ment value(s). See Figure 3Ć3.
WM 0.000E0
XYZ 0.000E0,0.000E0,0.000E0
Measurement
Units
(See Table 3Ć3)
Measurement
Value
Exponential
Multiplier
Figure 3Ć3:ăRSĆ232D Report Format
Measurement unit abbreviations are listed in TableĂ3Ć3.
Values are displayed with four significant digits, and a
oneĆdigit exponential multiplier with an optional minus
sign. For multiple values (for example, the three values
reported for an XYZ reading), commas separate the
values.
Each line is terminated by a carriage return and line
feed (CR/LF).