Tektronix CSA8000B Uživatelský manuál Strana 275

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 354
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 274
Appendix B: Automatic Measurements Reference
CSA8000B & TDS8000B User Manual
B-13
Table B- 2: Pulse Measurements - Timing (cont.)
Name Definition
+Width The horizontal interval between the crossings of the rising and falling edges at the mid-refer-
ence level of the f irst posi tive pulse i n the measurement region.
+Width = Tcross2 Tcross1
Where Tcross1 and Tcross2 are t he t wo consecutive horizontal crossings on the first
positive pulse.
The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.
See Measurement Reference Parameters and Methods on page B--56 or in the online help. Also
see Reference Levels Method on page 3--79.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a M easurem ent on page 3--83.
-Width The horizontal interval between the crossings of the falling and rising edges at the mid-refer-
ence level of the f irst negat ive pulse in t he measurement region.
--Width = Tcross2 Tcross1
Where Tcross2 and Tcross1 are t he t wo consecutive horizontal crossings on the first
negative pul se.
The mid-reference level is adjustable and defaults to 50% of the pulse amplitude.
If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).
Zobrazit stránku 274
1 2 ... 270 271 272 273 274 275 276 277 278 279 280 ... 353 354

Komentáře k této Příručce

Žádné komentáře