Tektronix AWG710B Uživatelský manuál Strana 451

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 580
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 450
Appendix B: Performance Verification (AWG710B)
AWG710&AWG710B Arbitrary Waveform Generator User Manual B-19
Amplitude and Offset Accuracy Tests (Normal Out), (except option 02)
These procedures check the accuracy of the amplitude and offset outputs of the
AWG710B Arbitrary Waveform Generator.
Check Amplitude
Accuracy
Do the following steps to install the test hookup and set the test equipment controls:
1. Use a 50 BNC coaxial cable, a SMA(Ma)–BNC(Fe) adapter, a 50
precision terminator, and a BNC–to–dual banana adapter to connect the
AWG710B Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B-9).
Figure B-9: Amplitude accuracy initial test hookup
NOTE. The amplitude and offset accuracy checks are structured as a continuous
test. The next test uses the control settings from the previous test and uses the next
step in the sequence file.
NOTE. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT (CH1
or CH1) LED is off.
If the other OUTPUT LED is on, push the CH1
or CH1 OUT button to turn off the
output.
Equipment
required
A 50 BNC coaxial cable, a 50 precision terminator, a
SMA(Ma)–BNC(Fe) adapters, BNC (female)–to–dual banana adapter,
and a digital multimeter (DMM).
Prerequisites
The AWG710B Arbitrary Waveform Generator must meet the
prerequisites listed on page B-7.
AWG710B Arbitrary Waveform Generator
DMM
50 BNC coaxial cable
CH1 output connector
+ SMA(Ma)-BNC(Fe) adapter
+ BNC 50 precision terminator
+ 50 BNC coaxial cable
Input connector
+ BNC-to-dual banana adapter
+ 50 BNC coaxial cable
Zobrazit stránku 450
1 2 ... 446 447 448 449 450 451 452 453 454 455 456 ... 579 580

Komentáře k této Příručce

Žádné komentáře