Commands Listed in Alphabetical Order
Table 2-37: DATa and WFMOutpre Parameter Settings (cont.)
WFMOutpre Settings
DATa:ENCdg
Setting :ENCdg :BN_Fmt :BYT_Or :BYT_NR
RPBinary BIN RP
MSB
1,2
SRIbinary
BIN RI
LSB
1,2
SRPbinary
BIN RP
LSB
1,2
Examples
DATA:ENCDG? might return :DATa:ENCDG SRPB INARY for the format of the
outgoing w aveform data .
DATA:ENCDG RPBinary sets the data encoding format to be a positive integer
where the most signicant byte is transferred rst.
DATa:SOUrce
Sets or returns the location o f the waveform data transferred from the oscilloscope
by the CURVe?query.
Group
Waveform Transfer
Syntax
DATa:SOUrce
{CH1|CH2|CH3|CH4|MATH|REF1 |REF2|REF3|REF4|D0|D1|D2|D3|D4|D5|D6|D7|D8
DATa:SOUrce?
Related Commands
CURVe
Arguments
CH1–CH4 species which analog channel data will be transferred from the
oscilloscope to the controller, channels 1 through 4.
MATH species that the Math waveform data will be transferred from the
oscilloscope to the controller.
REF1–REF4 species which Reference wave form data will be transferred from
the oscilloscope to the controller, waveforms, 1 through 4.
D0–D15 species which digital channel data will be transferred from t he
oscilloscope to the controller, channels 0 through 15. (MSO models only)
DIGital species that the Digital waveform data w ill be transferred from the
oscilloscope to the controller. (MSO models only)