Commands Listed in Alphabetical Order
MEASUrement:
REFLevel:ABSolute:MID[1]
Sets or returns the mid reference level, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to Absolute. This command affects the
results of pe
riod, frequency, delay, and all cyclic measurements.
NOTE. this command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSol ute:MID[1] <NR3>
MEASUrement:REFLevel:ABSol ute:MID[1]?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is the mid reference level, in volts. The default is 0.0 V.
Examples
MEASUREMENT:REFLEVEL:ABSOL UTE:MID 1 .71 sets the mid reference level
to .71 V.
MEASUREMENT:REFLEVEL:ABSOL UTE:MID? might return
:MEASUREMENT:REFLEVEL:ABSO LUTE:MID 0.7100E+00 indicating
that the absolute mid1 reference level is set to .71 V.
MEASUrement:REFLevel:ABSolute:MID2
Sets or returns the mid reference level for the "to" waveform when
taking a delay measurement, and is the 50% reference level when
MEASUrement:REFLevel:METHod is set to Absolute. This command affects the
results of delay measurements.
NOTE. this command affects the associated reference level parameter for all
MEASurements:IMMed and the four periodic measurements.
Group
Measurement
Syntax
MEASUrement:REFLevel:ABSol ute:MID2 <NR3>
MEASUrement:REFLevel:ABSol ute:MID2?
MSO4000 and DPO4000 Series Programmer Manual 2-229