Tektronix DPO4000 Uživatelský manuál Strana 278

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 585
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 277
Commands Listed in Alphabetical Order
Conditions
This command re
quires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:SOURce {VOLTage|CURRen t}
POWer:HARMonics:SOURce?
Arguments
VOLTage species voltage source waveform for harmonic tests.
CURRent species current source waveform for harmonic tes ts.
Examples
POWER:HARMONICS:SOURCE VOLTage sets voltage source for harmonic tests.
POWER:HARMONICS:SOURCE? might r eturn :POWer:HARMonics:SO URCE
CURRENT
indicating that the harmonic source is set to current.
POWer:HARMonics:STANDard
Sets or returns the standard for harmonics tests.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:HARMonics:STANDard {NONe|IEC|MIL }
POWer:HARMonics:STANDard?
Arguments
NONe sets no standard for harmonic tests.
IEC sets IEC 610003-2 standard for harmonic tests.
MIL sets MIL1399 standard for harmonic tests.
Examples
POWER:HARMONICS:STANDARD IEC sets IEC standard for harmonic tes ts.
POWER:HARMONICS:STANDARD? might return :POWer:HARMonics:STANDard
NONe
indicating that no standard is set.
POWer:INDICators
Sets or returns the state of the measurement indicators for the power application.
2-268 MSO4000 and DPO4000 Series Programmer Manual