Commands Listed in Alphabetical Order
Arguments
<NR1> is the num
ber of samples used for the mean and standard deviation
statistical accumulations
Examples
POWER:STATISTICS:WEIGHTIN G 4 sets statistical weighting to four samples.
POWER:STATISTICS:WEIGHTIN G ? might return
:POWER:STATISTICS:WEIGHTI NG 4 indicating that s tatistics weighting is
currently set to 4 samples.
POWer:SWLoss:CONDCALCmethod
Sets or returns the power application switching loss conduction calculation
method.
Conditions
This command requires a DPO4PWR application module.
Group
Power
Syntax
POWer:SWLoss:CONDCALCmeth od {VOLTage|RDSon|VCEsat}
POWer:SWLoss:CONDCALCmeth od?
Arguments
VOLTage se
ts voltage as the conduction calculation method.
RDSon sets RDSon as the conduction calculation method.
VCEsat sets VCEsat as the conduction calculation method.
Examples
POWER:SWLOSS:CONDCALCMETH OD RDSon sets the conduction calculation
method to RDSon.
POWER:SWLOSS:CONDCALCMETH OD ? might return
:POWER:SWLOSS:CONDCALCMET HOD VOLT indicating that the conduction
calculation method is set to Voltage.
POWer:SWLoss:CONDuction:ENERGY:MAX? (Query Only)
Returns the maximum conduction energy for switching loss calculation.
Conditions
This command requires a DPO4PWR application module.
Group
Power
2-298 MSO4000 and DPO4000 Series Programmer Manual