Commands Listed in Alphabetical Order
Syntax
DIAg:LOOP:OPT
ion:NTIMes <NR1>
DIAg:LOOP:OPTion:NTIMes?
Arguments
<NR1> is the number of self-test loops.
Examples
DIAG:LOOP:OPTION:NTIMES 3 sets the self-test loop to run three times.
DIAG:LOOP:OPTION:NTIMES? might return :DIAG:LOOP:OPTION:NTIMES 5,
indicating the self-test loop is set to run ve times.
DIAg:LOOP:STOP (No Query Form)
Stops the self-test at the end of the current loop.
Group
Calibration and Diagnostic
Syntax
DIAg:LOOP:STOP
Examples
DIAG:LOOP:STOP stops the self test at the end of the current loop.
DIAg:RESUlt:FLAg? (Query Only)
Returns the pass/fail status from the last self-test sequence exec ution. Use this
query to determine which test(s) has failed.
Group
Calibration and Diagnostic
Syntax
DIAg:RESUlt:FLAg?
Related Commands
DIAg:RESUlt:LOG?
Returns
PASS indicates that all of the selected self (diagnostic) tests have passed.
FAIL indicates that at leas t one of the selected self (diagnostic) tests has failed.
IN PROG RESS indicates t hat at least one of the selected self (diagnostic) tests is
ongoing.
NOT RUN indicates that the selected diagnostic test suite has not been run since
the instrument was powered on.
MSO4000 and DPO4000 Series Programmer Manual 2-155