Commands Listed in Alphabetical Order
the d istance fr
om DATa:STARt to
DATa:STOP stays smaller than the increa sed
record length.
Group
Waveform Transfer
Syntax
DATa:STOP <NR1>
DATa:STOP?
Related Commands
CURVe, DATa, DATa:STARt, WFMInpre:NR_Pt, WFMOutpre:NR_Pt?
Arguments
<NR1> is the last data point that will be transferred, which ranges from 1 to
the record length. If
<NR1> is greater than the record length, then data will be
transferred up to the record length. If both DATa:STARt and DATa:STOP are
greater than the record length, the last data point in the record is returned.
DATa:STARt and DATa:STOP are order independent. When DATa:STOP is less
than DATa:STARt, the values will be swapped internally for the CURVE? query.
If you always want to transfer comple te waveforms, set DATa:STARt to 1 and
DATa:STOP to the maximum record length, or larger.
Examples
DATA:STOP? might return :DATA:S TOP 14900 indicating that 14900 is the last
waveform data point that will be transferred.
DATA:STOP 15000 species that the waveform transfer will stop at data point
15000.
DATE
Sets or returns the date the oscilloscope displays.
Group
Miscellaneous
Syntax
DATE <Q String>
DATE?
Related Commands
TIME
Arguments
<QString> is a date in the form "yyyy-mm-dd" where yyyy refers to a four-digit
year number, mm refers to a two-digit month number f rom 01 to 12, and dd refers
to a two-digit day number in the month.
MSO4000 and DPO4000 Series Programmer Manual 2-151