Tektronix DPO4000 Uživatelský manuál Strana 467

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 585
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 466
Commands Listed in Alphabetical Order
TRIGger:A:VI
Deo:CUSTom:LINEPeriod
Sets or queries the line period for the standard under test. Use this command only
when the video format is set to custom.
Conditions
This command requires a DPO4VID application module.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:LIN EPeriod <NR3>
TRIGger:A:VIDeo:CUSTom:LIN EPeriod?
Arguments
<NR3> is the custom video line period.
Examples
TRIGGER:A:VIDEO:CUSTOM:LIN EPERIOD 50.5E-6 sets the video line period
to 50.5 μs.
TRIGGER:A:VIDEO:CUSTOM:LIN EPERIOD? might return
TRIGGER:A:VIDEO:CUSTOM:LIN EPERIOD 63.5600E-6 indicating
the video line period is set to 63.56 μs.
TRIGger:A:VIDeo:CUSTom:SCAN
Sets or returns the horizontal line scan rate of the A v ideo trigger. Use this
command only when the video format is set to custom.
Conditions
This command requires a DPO4VID application module.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:SCA N
{RATE15K|RATE20K|RATE25K|R ATE35K|RATE50K}
TRIGger:A:VIDeo:CUSTom:SCA N?
Arguments
RATE15 sets the range of the video line scan rate to 15 kHz through 20 kHz. This
is the standard broadcast rate.
RATE20 sets the range of the video line scan rate to 20 kHz through 25 kHz.
RATE25 sets the range of the video line scan rate to 25 kHz through 35 kHz
RATE35 sets the range of the video line scan rate to 35 kHz through 50 kHz
MSO4000 and DPO4000 Series Programmer Manual 2-457
Zobrazit stránku 466
1 2 ... 462 463 464 465 466 467 468 469 470 471 472 ... 584 585